SPX-100/150/250B-Z biochemical incubator instruction manual

SPX biochemical incubator is a high-precision thermostat with cold and heat control. It can be used for plant culture, breeding test, culture, preservation of bacteria, mold, microorganisms, BOD measurement of water analysis and constant temperature test for other purposes. It is an ideal test equipment for scientific research units such as biological genetic engineering, medical treatment, health and epidemic prevention, drug testing, agriculture, animal husbandry and aquatic products.

Structural characteristics of biochemical incubator The chassis is made of high-quality thin plate, and the surface is painted firmly and beautifully. A double-glazed glass window is placed in the middle of the box door, and the interior of the box is illuminated to make the test items clear at a glance. The inner liner is made of mirror stainless steel sheet and is made of four-corner arc shape. The temperature inside the box is uniform, durable and easy to clean. The temperature control part adopts a new intelligent digital display temperature regulator with single chip technology. Users can set and adjust the temperature and time by operating the touch panel of the control panel according to different requirements to achieve the test purpose.

Working principle of biochemical incubator The incubator is converted into an electrical signal by the actual temperature felt by the temperature sensor located in the box, and the microcomputer or the refrigeration compressor is controlled by the microcomputer to reach the required temperature.

Biochemical incubator technical parameters
1. Nominal volume: 100L, 150L, ​​250L;
2, temperature control range: 5 ~ 50 ° C;
3, temperature fluctuation tolerance: ± 1 ° C (100L, 150L) ± 1.5 ° C (250L);
4, temperature uniformity tolerance: ± 1.5 ° C (100L, 150L) ± 2 ° C (250L);
5, the power supply voltage: AC 220V / 50Hz;
6, input power: 280W (100L) 320W (150L, ​​250L);
7. Working environment: the ambient temperature is 10~30°C and the relative humidity is below 70%;
8, refrigerant: R12;
9, equipment category: Class I B type.

Biochemical incubator installation
1. This equipment should be installed in a room that is away from direct sunlight and is ventilated and dry. The equipment and the wall must have a distance of more than 10CM.
2. The cabinet can be adjusted with the bottom adjustment screw to make the cabinet stable.
3. This equipment uses AC 220V/50Hz power supply, and there must be a reliable grounding wire in the power supply circuit to ensure safe use.
1. Turn on the power, turn on the power switch, and make the switch in the “on” position, and the device enters the power-on state. At this point, the temperature and time setting operation can be started.
2. The temperature/time button can be used to switch between temperature and time:
1) Temperature setting: Press the temperature/time button and the temperature setting indicator of the meter will light. Press the Set button to enter the SV state, set one of the three numbers in the display window to flash, then press the shift button ( < ) to shift the number of flashes to the desired setting, then press the ∨ or ∧ button to change the temperature setting. Set to the desired temperature (minimum 0 °C), and finally press the Set button to save the set value to the microcontroller. At this time, the left display window displays the measured temperature, and the right display window displays the set temperature.
2) Time setting: Press the temperature/time button, the meter's time setting indicator lights. Press the Set button to enter the TV state, set one of the three numbers in the display window to flash, then press the shift button ( < ) to shift the flashing number to the desired setting, then press the ∨ or ∧ button to change the time setting. Set the value until the required time (minimum 1h), and finally press the set button to save the set value to the microcontroller. At this time, the left display window displays the measurement time, and the right display window displays the set time.
Timing function: When the instrument is powered, the timing function starts. After the timing is over, the output is turned off. The heater and the refrigeration compressor stop working, and the temperature inside the incubator will gradually reach the ambient temperature.
3. Status indication: After the setting is completed, the device runs according to the set parameters. When the measured temperature is lower than the set temperature, the device enters the heating state and the heating indicator lights. Conversely, when the measured temperature is higher than the set temperature, the cooling indicator lights up and the device enters the cooling state. When the measured temperature is higher than the set temperature by 4 °C, the alarm indicator lights and the buzzer sounds.
4. Parameter modification: If you want to modify the temperature and time parameters during operation, modify them according to procedures 1 and 2.
5. Quick setting: When setting parameters, one of the three numbers will flash. If you press the ∧ key, the last digit of the set value will be increased by one; if you press this button, the value will be set after about 3 seconds. Add 1 continuously; if you press the ∨ key, the set value is decreased by 1; if you press and hold the key, the set value is continuously decreased by 1 after about 3 seconds.

Biochemical incubator precautions and maintenance
1. During the handling of the incubator, it is forbidden to invert and lay flat more than 45 degrees.
2. Do not change the set value frequently during use to avoid overloading the compressor and affecting the service life of the device.
3. When there is no need for lighting in the box, the lighting switch should be placed in the "off" position to avoid affecting the upper layer temperature and prolong the life of the lamp.
4. This machine is equipped with two sets of fuses. If there is a malfunction during operation, please cut off the power supply first, check if the fuse is intact, and then check other parts.
5. In order to maintain the appearance of the device, do not wipe the surface with a corrosive solution. Wipe the cloth with a dry cloth or alcohol to keep the inside of the box clean.
6. When the equipment is not in use, keep the inside of the box dry and cut off the power.
7. To ensure uniform temperature inside the box, always check the normal operation of the axial fan in the box. During the experiment, the contents of the box should not be placed too densely and do not block the fan outlet to facilitate circulation of the airflow inside the box.
8. Do not touch, touch the temperature probe inside the box, causing the temperature to run out of control.
9. If the equipment fails, you should ask a professional to repair or contact the sales department of the factory. Users should not arbitrarily dismantle.
10. Our products implement “three guarantees”. From the date of product sale, if there are quality problems within one year, the factory is responsible for free maintenance (except for man-made damage). Exceeding the "three guarantees" period of discretionary repairs.